Print Email Facebook Twitter Electrophysiological Evaluation of the Continuity and Transmurality of Surgical Ablation Lesions in Ex Vivo Perfused Porcine Hearts by Unipolar High-Resolution Mapping Title Electrophysiological Evaluation of the Continuity and Transmurality of Surgical Ablation Lesions in Ex Vivo Perfused Porcine Hearts by Unipolar High-Resolution Mapping Author Wenink, Kristel (TU Delft Mechanical, Maritime and Materials Engineering) Contributor de Groot, N.M.S. (mentor) van Schie, M.S. (mentor) Taverne, Y.J.H.J. (graduation committee) Hendriks, R.C. (graduation committee) Degree granting institution Delft University of TechnologyErasmus Universiteit RotterdamUniversiteit Leiden Programme Technical Medicine Date 2023-03-20 Abstract Introduction: Currently, there is no method available for intra-operative evaluation of the completeness (transmurality and continuity) of surgical ablation lesions. This study aimed to investigate the changes in electrogram characteristics and activation patterns caused by different (non)transmural and (dis)continuous ablation lesions in ex vivo perfused porcine hearts. Methods: Donation after circulatory death porcine hearts ex vivo perfused in Langendorff mode were used to perform ablation experiments in a controlled setting. Three subsequent radiofrequency ablation lesions – with different degrees of transmurality and continuity - were created on the right ventricle using AtriCure’s Isolator Synergy Bipolar clamp. Electrograms of the lesion and surrounding tissue were recorded by unipolar high-resolution mapping. These measurements were executed during pacing perpendicular to the ablation lesion from two sides and during intrinsic cardiac rhythm. Electrograms were processed using custom-made software. The inter-electrode conduction time, potential voltage, potential slope, and R-to-S-amplitude ratio were analyzed.Results: The first radiofrequency application significantly affected all parameters in the lesion area. Conduction times increased, the potential voltage and slope decreased, and there was a loss of S-wave amplitude. The increase in conduction time and the decrease in voltage were less steep when there was a conduction gap in the ablation line. However, conduction time was less sensitive to lesion transmurality because it remained stable even when the lesion became more transmural. The potential voltage on the other hand, became significantly lower in transmural lesions, showing an overall decrease of 84% from baseline to the third (complete) lesion. The potential slope showed similar trends as the voltage, although it was less discriminative for (non)transmurality and (dis)continuity. The loss of the S-wave became significantly more pronounced with more radiofrequency delivery.Conclusions: Complete ablation lesions are characterized by a stable conduction time when applying subsequent ablation, a decrease in potential voltage of 84% on the lesion and its border zone, and loss of the contribution of S-wave amplitude. The combination of these parameters in one tool could help to detect incomplete surgical ablation lesions in the individual patient during Maze surgery. This could potentially reduce post-maze gap-related atrial tachyarrhythmias and thus improve long-term success rates. Subject Atrial fibrillationMaze surgerySurgical ablationEx vivo heart perfusionHigh-resolution mapping To reference this document use: http://resolver.tudelft.nl/uuid:f7aa4889-9b2b-4a8c-8db0-23981ab004b3 Part of collection Student theses Document type master thesis Rights © 2023 Kristel Wenink Files PDF MasterThesis_KristelWenink.pdf 4.02 MB Close viewer /islandora/object/uuid:f7aa4889-9b2b-4a8c-8db0-23981ab004b3/datastream/OBJ/view